The correct and incorrect way to calibrate a Compton suppression counting system for gamma-ray efficiency

Gamma-ray efficiency calculations for a germanium detector have been made for a Compton suppression system. Results have shown that for radionuclides that have gamma rays in coincidence the photopeaks can be severely depressed leading to erroneous results. While this can be overcome in routine neutr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of radioanalytical and nuclear chemistry 2013-04, Vol.296 (1), p.309-313
Hauptverfasser: Landsberger, S., Biegalski, S., Kapsimalis, R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Gamma-ray efficiency calculations for a germanium detector have been made for a Compton suppression system. Results have shown that for radionuclides that have gamma rays in coincidence the photopeaks can be severely depressed leading to erroneous results. While this can be overcome in routine neutron activation analysis using a comparator method, special consideration must be given to determine the suppression for coincident gamma rays when calculating the efficiency curve and for radionuclide activities. This is especially important for users of the k 0 method and for fission product identification using Compton suppression methods.
ISSN:0236-5731
1588-2780
DOI:10.1007/s10967-012-2123-1