Delayed migration of sapien valve following a transcatheter mitral valve‐in‐valve implantation

We report two cases of delayed migration of the Sapien XT device after a successful mitral valve‐in‐valve (VIV) implantation. The procedure was performed through a transapical approach. Echocardiography was used to choose the size of the Sapien XT device. Although the immediate results were satisfac...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Catheterization and cardiovascular interventions 2014-01, Vol.83 (1), p.E150-E154
Hauptverfasser: Bapat, Vinayak (Vinnie) Nilkanth, Khaliel, Feras, Ihleberg, Leo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report two cases of delayed migration of the Sapien XT device after a successful mitral valve‐in‐valve (VIV) implantation. The procedure was performed through a transapical approach. Echocardiography was used to choose the size of the Sapien XT device. Although the immediate results were satisfactory both the cases presented with severe regurgitation (1 week and 3 months). Investigations revealed atrial migration of the Sapien device in both the cases, which was confirmed at the time of reoperation. We discuss possible mechanisms, which could have resulted in the delayed migration and highlight the difference between VIV procedures in mitral position versus other positions. © 2013 Wiley Periodicals, Inc.