Comparison of instrumental methods for chlorinated dibenzo-p-dioxin (CDD) determination - interim results of a round-robin study involving GC-MS, MS-MS, and high resolution MS
A study was designed to determine the feasibility of using GC-MS-MS techniques for the determination of chlorinated dibenzo-p-dioxins (CDD) and dibenzofurans (CDF) in difficult environmental samples with reduced cleanup. Three fish and three sediment extracts were provided to six different laborator...
Gespeichert in:
Veröffentlicht in: | Chemosphere (Oxford) 1986, Vol.15 (9), p.1147-1156 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A study was designed to determine the feasibility of using GC-MS-MS techniques for the determination of chlorinated dibenzo-p-dioxins (CDD) and dibenzofurans (CDF) in difficult environmental samples with reduced cleanup. Three fish and three sediment extracts were provided to six different laboratories with GC-MS-MS capability for analysis. In addition, extracts were analysed by GC-low reduction MS (LRMS) and GC-high resolution MS (HRMS). At the time of this report, all participating GC-MS-MS laboratories had not completed their analysis of the supplied samples. All reporting laboratories found the analysis of these very complex samples difficult without extensive cleanup. GC-HRMS determination of some samples was possible with reduced cleanup compared to GC-LRMS. |
---|---|
ISSN: | 0045-6535 1879-1298 |
DOI: | 10.1016/0045-6535(86)90385-1 |