Evaluation of matrix-assisted laser desorption ionization-time-of-flight mass measurement accuracy by using delayed extraction
Mass measurement accuracy of a reflectron time-of-flight mass spectrometer equipped with delayed extraction is evaluated. Mass resolution of 10,000–15,000 is achieved routinely for peptides in the mass range of 1–6 ku ionized by matrix-assisted laser desorption ionization. The mass measurement accur...
Gespeichert in:
Veröffentlicht in: | Journal of the American Society for Mass Spectrometry 1996-10, Vol.7 (10), p.995-1001 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Mass measurement accuracy of a reflectron time-of-flight mass spectrometer equipped with delayed extraction is evaluated. Mass resolution of 10,000–15,000 is achieved routinely for peptides in the mass range of 1–6 ku ionized by matrix-assisted laser desorption ionization. The mass measurement accuracy of peptides with molecular weights of 1–4 ku is 10–15 ppm by using external calibration and better than 5 ppm by using internal calibration. The mass calibration remains accurate over a broad mass-to-charge ratio range (1–4 ku), even when the calibration curve is extrapolated several thousand mass units. |
---|---|
ISSN: | 1044-0305 1879-1123 |
DOI: | 10.1016/1044-0305(96)00027-X |