A laboratory comparison of the performance of the buddy lite™ and enFlow™ fluid warmers
Summary Maintenance of normothermia is crucial to avoid patient morbidity. Newly released fluid warming devices have become smaller in size, but this change might impair efficacy. We performed an evaluation of the buddy lite™ and enFlow™ fluid warmers. We measured inflow and outlet temperatures of t...
Gespeichert in:
Veröffentlicht in: | Anaesthesia 2013-11, Vol.68 (11), p.1161-1164 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Summary
Maintenance of normothermia is crucial to avoid patient morbidity. Newly released fluid warming devices have become smaller in size, but this change might impair efficacy. We performed an evaluation of the buddy lite™ and enFlow™ fluid warmers. We measured inflow and outlet temperatures of the devices at flow rates between 25 and 100 ml.min−1 using saline at room temperature or cooled to 10 °C. At a flow rate of 25 ml.min−1, the outlet temperature of the buddy lite was significantly higher than that of the enFlow (p |
---|---|