Trauma in Patients With Continuous-Flow Left Ventricular Assist Devices

Trauma-related failure of a continuous-flow left ventricular assist device (LVAD) has not previously been reported. We present 4 cases in which LVAD complications were likely caused by external trauma and led to failure of a HeartMate II device. In 1 case, the onset of symptoms was delayed and the p...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The American journal of cardiology 2013-11, Vol.112 (9), p.1520-1522
Hauptverfasser: Sarsam, Sinan H., MD, Meyers, Deborah E., MD, Civitello, Andrew B., MD, Agunanne, Enoch E., MD, Odegaard, Peggy, RN, BSN, Cohn, William E., MD, Frazier, O.H., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Trauma-related failure of a continuous-flow left ventricular assist device (LVAD) has not previously been reported. We present 4 cases in which LVAD complications were likely caused by external trauma and led to failure of a HeartMate II device. In 1 case, the onset of symptoms was delayed and the patient did not seek medical attention until months after the traumatic event. All 4 patients required surgical intervention, and 1 patient died of respiratory complications several months postoperatively. In conclusion, a history of external trauma should be considered as a possible etiologic factor when LVAD-supported patients in previously stable condition present with device malfunction.