Safety and Efficacy of a Totally Subcutaneous Implantable-Cardioverter Defibrillator
BACKGROUND—The most frequent complications associated with implantable cardioverter-defibrillators (ICDs) involve the transvenous leads. A subcutaneous implantable cardioverter-defibrillator (S-ICD) has been developed as an alternative system. This study evaluated the safety and effectiveness of the...
Gespeichert in:
Veröffentlicht in: | Circulation (New York, N.Y.) N.Y.), 2013-08, Vol.128 (9), p.944-953 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | BACKGROUND—The most frequent complications associated with implantable cardioverter-defibrillators (ICDs) involve the transvenous leads. A subcutaneous implantable cardioverter-defibrillator (S-ICD) has been developed as an alternative system. This study evaluated the safety and effectiveness of the S-ICD System (Cameron Health/Boston Scientific) for the treatment of life-threatening ventricular arrhythmias (ventricular tachycardia/ventricular fibrillation).
METHODS AND RESULTS—This prospective, nonrandomized, multicenter trial included adult patients with a standard indication for an ICD, who neither required pacing nor had documented pace-terminable ventricular tachycardia. The primary safety end point was the 180-day S-ICD System complication-free rate compared with a prespecified performance goal of 79%. The primary effectiveness end point was the induced ventricular fibrillation conversion rate compared with a prespecified performance goal of 88%, with success defined as 2 consecutive ventricular fibrillation conversions of 4 attempts. Detection and conversion of spontaneous episodes were also evaluated. Device implantation was attempted in 321 of 330 enrolled patients, and 314 patients underwent successful implantation. The cohort was followed for a mean duration of 11 months. The study population was 74% male with a mean age of 52±16 years and mean left ventricular ejection fraction of 36±16%. A previous transvenous ICD had been implanted in 13%. Both primary end points were metThe 180-day system complication-free rate was 99%, and sensitivity analysis of the acute ventricular fibrillation conversion rate was >90% in the entire cohort. There were 38 discrete spontaneous episodes of ventricular tachycardia/ventricular fibrillation recorded in 21 patients (6.7%), all of which successfully converted. Forty-one patients (13.1%) received an inappropriate shock.
CONCLUSIONS—The findings support the efficacy and safety of the S-ICD System for the treatment of life-threatening ventricular arrhythmias.
CLINICAL TRIAL REGISTRATION—URLhttp://www.clinicaltrials.gov. Unique identifierNCT01064076. |
---|---|