Superparamagnetic nanoparticle detection using second harmonic of magnetization response
We introduce a method to improve the detection sensitivity for the magnetization M of superparamagnetic nanoparticles (MNP). The M response of MNP to an applied magnetic field H (M-H characteristics) could be divided into a linear region and a saturation region, which are separated at a transition p...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2013-09, Vol.84 (9), p.094702-094702 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We introduce a method to improve the detection sensitivity for the magnetization M of superparamagnetic nanoparticles (MNP). The M response of MNP to an applied magnetic field H (M-H characteristics) could be divided into a linear region and a saturation region, which are separated at a transition point H(k). When applying an excitation magnetic field (H(ac)) with a frequency ω0 and an additional dc bias field H(dc) = H(k), the second harmonic of M reaches the maximum due to the nonlinearity of the M-H characteristics. It is stronger than any other harmonics and responsible for small H(ac) without a threshold. The second harmonic selected as the readout criterion for M response of MNP is systematically analyzed and experimentally proven. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.4820447 |