Balloon-assisted over-the-wire technique for placement of the venous outflow component of the Hemodialysis Reliable Outflow (HeRO) device

A modified technique for placement of the venous outflow component (VOC) of the Hemodialysis Reliable Outflow (HeRO) device (Hemosphere Inc, Minneapolis, Minn) is described. The purpose of the technique is to improve the system's trackability and facilitate device insertion in patients with cen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of vascular surgery 2013-10, Vol.58 (4), p.1117-1119
Hauptverfasser: Ferral, Hector, MD, Behrens, George, MD, Tumer, Yanki, MD, Gueyikian, Sebouh A., MD, Hollinger, Edward F., MD, Jensik, Stephen C., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A modified technique for placement of the venous outflow component (VOC) of the Hemodialysis Reliable Outflow (HeRO) device (Hemosphere Inc, Minneapolis, Minn) is described. The purpose of the technique is to improve the system's trackability and facilitate device insertion in patients with central venous occlusion. Device preparation requires placement of a 6-mm × 4-cm angioplasty balloon within the leading end of the VOC. The leading 2 cm of the balloon are placed just distal to the radiopaque marker of the VOC. The balloon is inflated to profile and locked in this position within the leading end of the VOC. The VOC and balloon combination is advanced over the wire through the 20F peel-away sheath provided by the manufacturer. The described technique was used to successfully implant the HeRO device in 12 patients with central venous occlusion. This technique is recommended for placement of the VOC of the HeRO device in patients with central venous occlusions.
ISSN:0741-5214