Automated Measurement of Aqueous Aluminum by the Pyrocatechol Violet Method
In response to our need for a rapid method of analyzing numerous aluminum samples, we developed an automated procedure for measuring total monomeric aluminum in aqueous samples. Two separate designs for a continuous microflow analyzer are presented which enable detection over a combined range of 20...
Gespeichert in:
Veröffentlicht in: | International journal of environmental analytical chemistry 1993-01, Vol.50 (3), p.173-182 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In response to our need for a rapid method of analyzing numerous aluminum samples, we developed an automated procedure for measuring total monomeric aluminum in aqueous samples. Two separate designs for a continuous microflow analyzer are presented which enable detection over a combined range of 20 μg Al/l to 1000 μg Al/l. The best fit to the calibration curve for the high sensitivity manifold is not linear (r
2
= 0.998) as previously reported but third order (cubic, r
2
= 1.000). The choice of equations for curve fit for the low sensitivity range is less critical as first and third order equations provide comparable fit. Between-sample contamination is less that 3% and reproducibility is high; less than 1% variation for high (200 - 1000 μg Al/l) concentrations and less than 7% for lower (20 - 300 μg Al/l concentrations. Analysis of both water and soil extract samples for aluminum should be facilitated by use of this system. |
---|---|
ISSN: | 0306-7319 1029-0397 |
DOI: | 10.1080/03067319308027595 |