Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions
Purpose To determine the value of a metal artefact reduction (MAR) algorithm with iterative reconstructions for dental hardware in carotid CT angiography. Methods Twenty-four patients (six of which were women; mean age 70 ± 12 years) with dental hardware undergoing carotid CT angiography were includ...
Gespeichert in:
Veröffentlicht in: | European radiology 2013-10, Vol.23 (10), p.2687-2694 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To determine the value of a metal artefact reduction (MAR) algorithm with iterative reconstructions for dental hardware in carotid CT angiography.
Methods
Twenty-four patients (six of which were women; mean age 70 ± 12 years) with dental hardware undergoing carotid CT angiography were included. Datasets were reconstructed with filtered back projection (FBP) and using a MAR algorithm employing normalisation and an iterative frequency-split (IFS) approach. Three blinded, independent readers measured CT attenuation values and evaluated image quality and degrees of artefacts using axial images, multi-planar reformations (MPRs) and maximal intensity projections (MIP) of the carotid arteries.
Results
CT attenuation values of the internal carotid artery on images with metal artefacts were significantly higher in FBP (324 ± 104HU) datasets compared with those reconstructed with IFS (278 ± 114HU;
P
|
---|---|
ISSN: | 0938-7994 1432-1084 |
DOI: | 10.1007/s00330-013-2885-z |