X-ray bang-time measurements at the National Ignition Facility using a diamond detector

A chemical vapor deposition polycrystalline photoconductive diamond detector was fielded at NIF to measure the time of peak x-ray emission, or x-ray bang time, of inertial confinement fusion implosions. Imaging the capsule with a pinhole provides contrast against Hohlraum emission, allowing clear id...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2012-10, Vol.83 (10), p.10E105-10E105
Hauptverfasser: Barrios, M A, MacPhee, A, Regan, S P, Kimbrough, J, Nagel, S R, Benedetti, L R, Khan, S F, Bradley, D, Bell, P, Edgell, D, Collins, G W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A chemical vapor deposition polycrystalline photoconductive diamond detector was fielded at NIF to measure the time of peak x-ray emission, or x-ray bang time, of inertial confinement fusion implosions. Imaging the capsule with a pinhole provides contrast against Hohlraum emission, allowing clear identification of the capsule component in the raw scope trace. X-ray bang time was measured to within ±41-46 ps with the internal photoconductive diamond detector.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.4729667