Effects due to a Pu-C source on a HPGe detector and the corresponding neutron shielding
A gamma spectrum of a Pu-C source is measured using a p-type HPGe detector, whose three peaks (full energy, single-escape and double-escape peak) can be used as a calibration source for the beam energy measurement system of BEPCII. The effect of fast neutron damage on the energy resolution of the HP...
Gespeichert in:
Veröffentlicht in: | Chinese physics C 2011-07, Vol.35 (7), p.660-667 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A gamma spectrum of a Pu-C source is measured using a p-type HPGe detector, whose three peaks (full energy, single-escape and double-escape peak) can be used as a calibration source for the beam energy measurement system of BEPCII. The effect of fast neutron damage on the energy resolution of the HPGe detector is studied, which indicates that the energy resolution begins to deteriorate when the detector is subject to 2×107 n/cm^2 fast neutrons. The neutron damage mechanism and detector repair methods are reviewed. The Monte Carlo simulation technique is utilized to study the shielding of the HPGe detector from the fast neutron radiation damage, which is of great significance for the future commissioning of the beam energy measurement system. |
---|---|