Performance of the Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry system for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories
We evaluated the performance of the Vitek MS for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. With a total of 424 well-characterized isolates, the results of the Vitek MS were compared to those of conventional methods and 16S rRNA gene sequencing. T...
Gespeichert in:
Veröffentlicht in: | Journal of medical microbiology 2013-09, Vol.62 (9), p.1301-1306 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated the performance of the Vitek MS for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. With a total of 424 well-characterized isolates, the results of the Vitek MS were compared to those of conventional methods and 16S rRNA gene sequencing. The Vitek MS correctly identified 97.9 % of the isolates tested to species level. The Vitek MS correctly identified the species of 97.2 % of the staphylococci (95.9 % of coagulase-negative staphylococci), 97.8 % of the streptococci, and 100 % of the enterococci. For the identification of Gram-positive cocci isolates, the overall concordance rate between conventional identification and the Vitek MS was 94.5 %. The Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system can be a reliable and rapid method for the identification of most relevant Gram-positive cocci. In addition, expanding the database of the Vitek MS, especially for coagulase-negative staphylococci, is needed to enhance the performance of the Vitek MS. |
---|---|
ISSN: | 0022-2615 1473-5644 |
DOI: | 10.1099/jmm.0.062950-0 |