Phototherapy Device Effectiveness in Nigeria: Irradiance Assessment and Potential for Improvement

This study investigated the effectiveness of simple-to-implement adjustments of phototherapy devices on irradiance levels in a cross-section of Nigerian hospitals. A total of 76 phototherapy devices were evaluated in 16 hospitals while adjustments were implemented for a subset of 25 devices for whic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of tropical pediatrics (1980) 2013-08, Vol.59 (4), p.321-325
Hauptverfasser: CLINE, Benjamin K, VREMAN, Hendrik J, FABER, Kelly, LOU, Hannah, DONALDSON, Krista M, AMUABUNOSI, Emmanuel, OFOVWE, Gabriel, BHUTANI, Vinod K, OLUSANYA, Bolajoko O, SLUSHER, Tina M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study investigated the effectiveness of simple-to-implement adjustments of phototherapy devices on irradiance levels in a cross-section of Nigerian hospitals. A total of 76 phototherapy devices were evaluated in 16 hospitals while adjustments were implemented for a subset of 25 devices for which consent was obtained. The mean irradiance level was 7.6 ± 5.9 µW/cm(2)/nm for all devices prior to adjustments. The average irradiance level improved from 9.0 µW/cm(2)/nm to 27.3 µW/cm(2)/nm for the adjusted group (n = 25) compared with 6.8 ± 5.4 µW/cm(2)/nm for the unadjusted group (n = 51). Simple, inexpensive adjustments to phototherapy devices with sub-optimal irradiance levels can significantly improve their effectiveness to acceptable international standards and should be widely promoted in resource-constrained settings.
ISSN:0142-6338
1465-3664
DOI:10.1093/tropej/fmt027