Development of a sandwich format, amperometric screen-printed uric acid biosensor for urine analysis

A screen-printed carbon electrode (SPCE) incorporating the electrocatalyst cobalt phthalocyanine (CoPC), fabricated using a water-based ink formulation, has been investigated as the base transducer for a uric acid biosensor. A sandwich biosensor was fabricated by first depositing cellulose acetate (...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical biochemistry 2012-09, Vol.428 (1), p.39-43
Hauptverfasser: Kanyong, P., Pemberton, R.M., Jackson, S.K., Hart, J.P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A screen-printed carbon electrode (SPCE) incorporating the electrocatalyst cobalt phthalocyanine (CoPC), fabricated using a water-based ink formulation, has been investigated as the base transducer for a uric acid biosensor. A sandwich biosensor was fabricated by first depositing cellulose acetate (CA) onto this transducer (CoPC–SPCE), followed by uricase (UOX) and finally a polycarbonate (PC) membrane; this device is designated PC–UOX–CA–CoPC–SPCE. This biosensor was used in conjunction with chronoamperometry to optimize the conditions for the analysis of urine: temperature, 35°C; buffer, pH 9.2; ionic strength, 50mM; uricase, 0.6U; incubation time, 180s. The proposed biosensor was applied to urine from a healthy subject. The precision determined on unspiked urine (n=6) was 5.82%. Urine was fortified with 0.225mM UA, and the resulting precision and recovery were 4.21 and 97.3%, respectively. The linear working range of the biosensor was found to be 0.015 to 0.25mM (the former represents the detection limit), and the sensitivity was calculated to be 2.10μA/mM.
ISSN:0003-2697
1096-0309
DOI:10.1016/j.ab.2012.05.027