Instrument for gas permeation measurements at high pressure and high temperature
An instrument was built for the permeation testing of flat polymer membranes under pressures up to 3.0 MPa and temperatures up to 300 °C. The high pressure, high temperature cell uses aluminum tape and a graphite gasket to minimize the leak from the high pressure side to the low pressure side, makin...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2013-06, Vol.84 (6), p.065107-065107 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An instrument was built for the permeation testing of flat polymer membranes under pressures up to 3.0 MPa and temperatures up to 300 °C. The high pressure, high temperature cell uses aluminum tape and a graphite gasket to minimize the leak from the high pressure side to the low pressure side, making possible the permeability measurements of slow diffusing gases such as N2. A computer program developed on a LabVIEW platform fully controls the instrument and data acquisition. It incorporates algorithms to automatically adjust the downstream volume, repressurize the upstream volume, vent the downstream volume to prevent over pressurization, and change the temperature of the permeation cell. The percent relative standard deviation of the permeability measurements was |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.4808285 |