Determination of coplanar polychlorinated biphenyl (CB) congeners in seal tissues by chromatography on active carbon, dual-column high resolution GC/ECD and high resolution GC/high resolution MS

A method for measurement of toxic coplanar chlorinated biphenyl (CB) congeners in seal tissues is described. Chromatography on Carbopack C efficiently separates the non-ortho substituted coplanar CB congeners CB-37/81/77/126 and CB-169 from the other (mono- and di-ortho substituted) coplanar CB cong...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chemosphere (Oxford) 1992, Vol.24 (9), p.1181-1196
Hauptverfasser: Storr-Hansen, E., Cederberg, T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method for measurement of toxic coplanar chlorinated biphenyl (CB) congeners in seal tissues is described. Chromatography on Carbopack C efficiently separates the non-ortho substituted coplanar CB congeners CB-37/81/77/126 and CB-169 from the other (mono- and di-ortho substituted) coplanar CB congeners. In total 31 specific CB congeners are determined by this analytical method. Samples are fractionated, and analyzed by either high resolution dual-column electron capture gas chromatography (HR-GC/ECD) or high resolution gas chromatography/high resolution mass spectrometry (HR-GC/HR-MS) with isotope dilution technique. The method is validated through standard addition experiments, and by use of recovery surrogate standards. The recovery is (85 ± 16%; N=78) for the ECD analyses. Corresponding figures for the isotope dilution GC/MS analyses are (79 ± 14 (CB-77); 80 ± 15 (CB-126); 83 ± 17 (CB-169); N=63). Detection limits were 0,1 ng/g for the HR-GC/ECD analysis (undiluted extracts), and 0,005 ng/g by the isotope dilution GC/MS analysis.
ISSN:0045-6535
1879-1298
DOI:10.1016/0045-6535(92)90045-S