Segmental flow injection with ion-selective electrodes for the determination of fluoride in water
A flow analyser with enhanced sampling frequency combining the features of flow-injection and continuous-flow analysis is described. Parameters of the analyser were optimized to ensure the highest rate of analysis allowed by the bubble-through flow cell with fluoride-selective electrodes. Fluoride w...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 1993-03, Vol.274 (1), p.125-128 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A flow analyser with enhanced sampling frequency combining the features of flow-injection and continuous-flow analysis is described. Parameters of the analyser were optimized to ensure the highest rate of analysis allowed by the bubble-through flow cell with fluoride-selective electrodes. Fluoride was determined in 0.1-ml water samples with a sampling rate as high as 720 h
−1 and a relative standard deviation better than 1%. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/0003-2670(93)80612-O |