Flow cell with double slope factor for potentiometric determination of fluoride at low concentrations

A simple double cell unit with 3 ion-selective membranes is described for continuous detection of fluoride at high throughput. The membranes were single crystal fluoride selective types composed principally of lanthanum fluoride; the cell was made from PTFE tubes and Plexiglas. The internal referenc...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analyst (London) 1993, Vol.118 (7), p.859-861
Hauptverfasser: BORZITSKY, J. A, DVININ, A, PETRUKHIN, O. M, URUSOV, Y. I
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A simple double cell unit with 3 ion-selective membranes is described for continuous detection of fluoride at high throughput. The membranes were single crystal fluoride selective types composed principally of lanthanum fluoride; the cell was made from PTFE tubes and Plexiglas. The internal reference solution contained sodium fluoride and chloride with silver-silver chloride inner reference electrodes. The analyte solution, divided into 2 streams, was buffered before entering the cell in parallel with 2 streams of standard fluoride solution. This arrangement increased the slope factor and permitted a sampling rate of 120 per h. The relative standard deviation for natural waters whose fluoride concentrations were below 0.0001 mol per litre was 0.01-0.03.
ISSN:0003-2654
1364-5528
DOI:10.1039/AN9931800859