Validation of novel calibration scheme with traceable point-like (22)Na sources on six types of PET scanners
To improve the reliability and convenience of the calibration procedure of positron emission tomography (PET) scanners, we have been developing a novel calibration path based on traceable point-like sources. When using (22)Na sources, special care should be taken to avoid the effects of 1.275-MeV γ...
Gespeichert in:
Veröffentlicht in: | Annals of nuclear medicine 2013-05, Vol.27 (4), p.346-354 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To improve the reliability and convenience of the calibration procedure of positron emission tomography (PET) scanners, we have been developing a novel calibration path based on traceable point-like sources. When using (22)Na sources, special care should be taken to avoid the effects of 1.275-MeV γ rays accompanying β (+) decays. The purpose of this study is to validate this new calibration scheme with traceable point-like (22)Na sources on various types of PET scanners.
Traceable point-like (22)Na sources with a spherical absorber design that assures uniform angular distribution of the emitted annihilation photons were used. The tested PET scanners included a clinical whole-body PET scanner, four types of clinical PET/CT scanners from different manufacturers, and a small-animal PET scanner. The region of interest (ROI) diameter dependence of ROI values was represented with a fitting function, which was assumed to consist of a recovery part due to spatial resolution and a quadratic background part originating from the scattered γ rays.
The observed ROI radius dependence was well represented with the assumed fitting function (R (2) > 0.994). The calibration factors determined using the point-like sources were consistent with those by the standard cross-calibration method within an uncertainty of ±4 %, which was reasonable considering the uncertainty in the standard cross-calibration method.
This novel calibration scheme based on the use of traceable (22)Na point-like sources was successfully validated for six types of commercial PET scanners. |
---|---|