An experimental apparatus for simultaneously measuring Seebeck coefficient and electrical resistivity from 100 K to 600 K

In this paper, we report a fully automated experimental apparatus for measuring Seebeck coefficient and electrical resistivity of a sample simultaneously in a temperature range of 100-600 K. The Seebeck coefficient is measured using a quasi-steady temperature differential method in which two ceramic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2013-04, Vol.84 (4), p.043903-043903
Hauptverfasser: Guan, Aiqiang, Wang, Hanfu, Jin, Hao, Chu, Weiguo, Guo, Yanjun, Lu, Guiwu
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this paper, we report a fully automated experimental apparatus for measuring Seebeck coefficient and electrical resistivity of a sample simultaneously in a temperature range of 100-600 K. The Seebeck coefficient is measured using a quasi-steady temperature differential method in which two ceramic heaters are employed to alternately heat the sample. The sample holder is designed to reduce temperature disturbance on its base during a measurement cycle. To demonstrate the accuracy and reliability of the experimental setup, we have performed tests on reference materials including constantan and platinum.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.4798647