Cardiac output measured by uncalibrated arterial pressure waveform analysis by recently released software version 3.02 versus thermodilution in septic shock
To evaluate the 3.02 software version of the FloTrac/Vigileo™ system for estimation of cardiac output by uncalibrated arterial pressure waveform analysis, in septic shock. Nineteen consecutive patients in septic shock were studied. FloTrac/Vigileo™ measurements (COfv) were compared with pulmonary ar...
Gespeichert in:
Veröffentlicht in: | Journal of clinical monitoring and computing 2013-04, Vol.27 (2), p.171-177 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the 3.02 software version of the FloTrac/Vigileo™ system for estimation of cardiac output by uncalibrated arterial pressure waveform analysis, in septic shock. Nineteen consecutive patients in septic shock were studied. FloTrac/Vigileo™ measurements (COfv) were compared with pulmonary artery catheter thermodilution-derived cardiac output (COtd). The mean cardiac output was 7.7 L min
−1
and measurements correlated at r = 0.53 (
P
|
---|---|
ISSN: | 1387-1307 1573-2614 |
DOI: | 10.1007/s10877-012-9410-9 |