Portable optical-resolution photoacoustic microscopy with a pulsed laser diode excitation

Optical-resolution photoacoustic microscopy (OR-PAM) has been significantly improved in terms of spatial resolution, detection sensitivity, imaging speed, and penetration depth. However, the popular producibility of OR-PAM system is still limited by the size and cost of solid-state laser excitation....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied physics letters 2013-02, Vol.102 (5)
Hauptverfasser: Zeng, Lvming, Liu, Guodong, Yang, Diwu, Ji, Xuanrong
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Optical-resolution photoacoustic microscopy (OR-PAM) has been significantly improved in terms of spatial resolution, detection sensitivity, imaging speed, and penetration depth. However, the popular producibility of OR-PAM system is still limited by the size and cost of solid-state laser excitation. Here, we developed a portable laser-diode-based OR-PAM (LD-OR-PAM) system using a pulsed semiconductor laser source, which was operated at 905 ± 15 nm with a pulse energy as low as 4.9 μJ. The measured lateral resolution has been improved to ∼1.5 μm from hundreds of microns. The compact and inexpensive natures of LD-OR-PAM would promote the potential clinical applications such as in dermatology.
ISSN:0003-6951
1077-3118
DOI:10.1063/1.4791566