Stent flexibility versus concertina effect: Mechanism of an unpleasant trade-off in stent design and its implications for stent selection in the cath-lab

Abstract The “concertina effect”, longitudinal deformation of the proximal segments of a deployed stent when force is applied from a guide catheter or other equipment, is a recently recognised problem which seems to particularly affect more recent stent designs. Until now, flexibility and deliverabi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of cardiology 2013-04, Vol.164 (3), p.259-261
Hauptverfasser: Foin, N, Di Mario, C, Francis, D.P, Davies, J.E
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The “concertina effect”, longitudinal deformation of the proximal segments of a deployed stent when force is applied from a guide catheter or other equipment, is a recently recognised problem which seems to particularly affect more recent stent designs. Until now, flexibility and deliverability have been paramount aims in stent design. Developments have focussed on optimising these features which are commonly evaluated by clinicians and demanded by regulatory bodies. Contemporary stent designs now provide high flexibility by reducing the number of connecting links between stent segments and by allowing the connecting links to easily change their length. These design evolutions may, however, simultaneously reduce longitudinal strength and have the unintended effect of inducing some risk of longitudinal compression of the stent (the “concertina effect”) during difficult clinical cases. Progress in stent design and elimination of restenosis by drug coating has improved PCI outcome and enabled new applications. Here we discuss design trade-offs that shaped evolution and improvement in stent design, from early bare metal designs to the latest generation of drug eluting stent (DES) platforms. Longitudinal strength was not recognised as a critical parameter by clinicians or regulators until recently. Measurements, only now becoming publically available, seem to confirm vulnerability of some modern designs to longitudinal deformation. Clinicians could be more guarded in their assumption that changes in technology are beneficial in all clinical situations. Sometimes a silent trade-off may have taken place, adopting choices that are favourable for the vast majority of patients but exposing a few patients to unintended hazard.
ISSN:0167-5273