Bench comparative evaluation of a new generation and standard helmet for delivering non-invasive ventilation
Objective To evaluate the performance of a new helmet (NH) recently introduced into clinical use relative to that of the standard helmet (SH) in terms of delivering non-invasive continuous positive airway pressure (nCPAP) and pressure support ventilation (nPSV). Design This was a bench study using a...
Gespeichert in:
Veröffentlicht in: | Intensive care medicine 2013-04, Vol.39 (4), p.734-738 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
To evaluate the performance of a new helmet (NH) recently introduced into clinical use relative to that of the standard helmet (SH) in terms of delivering non-invasive continuous positive airway pressure (nCPAP) and pressure support ventilation (nPSV).
Design
This was a bench study using a mannequin connected to an active lung simulator. The SH was fastened to the mannequin by armpit braces, which are not needed to secure the NH.
Measurements
The inspiratory and expiratory variations in nCPAP delivered with two different simulated efforts (Pmus), were determined relative to the preset CPAP level. nPSV was applied at two simulated respiratory rates (RR) and two cycling-off flow thresholds. We measured inspiratory trigger delay (Delay
trinsp
), expiratory trigger delay (Delay
trexp
), time of synchrony (Time
sync
), trigger pressure drop (Δ
P
trigger
), airway pressure–time product during the triggering phase (PTPt), the initial 200 ms from the onset of the ventilator pressurization (PTP
200
), and the initial 300 and 500 ms from the onset of the simulated effort; this two latter parameters were expressed as the percentage of the area of ideal pressurization (PTP
300-index
and PTP
500-index
, respectively).
Results
In nCPAP, at both Pmus, the differences between the two interfaces at both Pmus were small and clinically irrelevant. In nPSV, regardless of the setting, NH resulted in significantly smaller trigger delays, Δ
P
trigger
, and PTPt. Time
sync
, PTP
200
, PTP
300-index
, and PTP
500-index
were also significantly higher with the NH compared to the SH, irrespective of the setting.
Conclusions
Compared to the SH, the NH is equally effective in delivering nCPAP and more effective in delivering nPSV, and it is used to avoid the need for armpit braces. |
---|---|