Radiation exposure during continuous and pulsed fluoroscopy
Abstract Background and Purpose: Previous studies using pulsed fluoroscopy have shown variable effects on radiation exposure because of the ramp and trail effect in older C-arm systems. This study compares radiation delivered in pulsed and continuous modalities using a modern C-arm system. Thermolum...
Gespeichert in:
Veröffentlicht in: | Journal of endourology 2013-03, Vol.27 (3), p.384-388 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Background and Purpose: Previous studies using pulsed fluoroscopy have shown variable effects on radiation exposure because of the ramp and trail effect in older C-arm systems. This study compares radiation delivered in pulsed and continuous modalities using a modern C-arm system.
Thermoluminescent dosimeters (TLDs) positioned in three body locations directly measured radiation dose during simulated ureteroscopy. Thirty pedal activations were administered using a pulsed or continuous mode to visualize an implanted guidewire and a radiopaque stone. TLD absorbed radiation and image quality were compared between imaging modes.
Pulsed fluoroscopy delivered less radiation compared with continuous fluoroscopy at each site: Anterior skin (0.10 vs 0.26 mGy, P |
---|---|