A comparative evaluation of a new automated assay for von Willebrand factor activity
Summary The ristocetin cofactor assay (VWF:RCo) is the reference method for assessing von Willebrand factor (VWF) activity in the diagnosis of von Willebrand's Disease (VWD). However, the assay suffers from poor reproducibility and sensitivity at low levels of VWF and is labour intensive. We ha...
Gespeichert in:
Veröffentlicht in: | Haemophilia : the official journal of the World Federation of Hemophilia 2013-03, Vol.19 (2), p.338-342 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Summary
The ristocetin cofactor assay (VWF:RCo) is the reference method for assessing von Willebrand factor (VWF) activity in the diagnosis of von Willebrand's Disease (VWD). However, the assay suffers from poor reproducibility and sensitivity at low levels of VWF and is labour intensive. We have undertaken an evaluation of a new immunoturbidimetric VWF activity (VWF:Ac) assay (INNOVANCE® VWF Ac. Siemens Healthcare Diagnostics, Marburg, Germany) relative to an established platelet‐based VWF:RCo method. Samples from 50 healthy normal subjects, 80 patients with VWD and 50 samples that exhibited ‘HIL’ (i.e. Haemolysis, Icterus or Lipaemia) were studied. VWF:Ac, VWF:RCo and VWF:Ag were performed on a CS–analyser (Sysmex UK Ltd, Milton Keynes, UK), all reagents were from Siemens Healthcare Diagnostics. The VWF:Ac assay, gave low intra‐ and inter‐assay imprecision (over a 31‐day period, n = 200 replicate readings) using commercial normal (Mean 96.2 IU dL−1, CV < 3.0%) and pathological (Mean 36.1 IU dL−1, CV |
---|---|
ISSN: | 1351-8216 1365-2516 |
DOI: | 10.1111/hae.12064 |