Failure of implantable cardioverter-defibrillator leads: A matter of lead size?

Background Small-diameter implantable cardioverter-defibrillator (ICD) Sprint Fidelis and Riata leads have been recalled owing to an increased risk of lead failure, thus arousing the suspicion that lead size might be a critical issue. Objective To compare the incidence of failure of small-diameter (...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Heart rhythm 2013-02, Vol.10 (2), p.184-190
Hauptverfasser: Rordorf, Roberto, MD, Poggio, Luca, MD, Savastano, Simone, MD, Vicentini, Alessandro, MD, Petracci, Barbara, MD, Chieffo, Enrico, MD, Klersy, Catherine, MD,MSc, Landolina, Maurizio, MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Small-diameter implantable cardioverter-defibrillator (ICD) Sprint Fidelis and Riata leads have been recalled owing to an increased risk of lead failure, thus arousing the suspicion that lead size might be a critical issue. Objective To compare the incidence of failure of small-diameter (≤8 F) and standard-diameter (>8 F) ICD leads implanted in a single center. Methods From January 2003 to December 2010, 190 Sprint Fidelis, 182 Riata/Riata ST, 99 Optim (Riata Optim/Durata), and 419 standard-diameter leads were implanted in our center. Results During a median follow-up of 33 months, the overall failure rate was 6.3%. Follow-up duration was similar for Sprint Fidelis, Riata, and standard-diameter leads but shorter for the Optim group. The failure rate was significantly higher in Sprint Fidelis leads than in both standard-diameter (4.8%/year vs 0.8%/year; P
ISSN:1547-5271