Failure of implantable cardioverter-defibrillator leads: A matter of lead size?
Background Small-diameter implantable cardioverter-defibrillator (ICD) Sprint Fidelis and Riata leads have been recalled owing to an increased risk of lead failure, thus arousing the suspicion that lead size might be a critical issue. Objective To compare the incidence of failure of small-diameter (...
Gespeichert in:
Veröffentlicht in: | Heart rhythm 2013-02, Vol.10 (2), p.184-190 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background Small-diameter implantable cardioverter-defibrillator (ICD) Sprint Fidelis and Riata leads have been recalled owing to an increased risk of lead failure, thus arousing the suspicion that lead size might be a critical issue. Objective To compare the incidence of failure of small-diameter (≤8 F) and standard-diameter (>8 F) ICD leads implanted in a single center. Methods From January 2003 to December 2010, 190 Sprint Fidelis, 182 Riata/Riata ST, 99 Optim (Riata Optim/Durata), and 419 standard-diameter leads were implanted in our center. Results During a median follow-up of 33 months, the overall failure rate was 6.3%. Follow-up duration was similar for Sprint Fidelis, Riata, and standard-diameter leads but shorter for the Optim group. The failure rate was significantly higher in Sprint Fidelis leads than in both standard-diameter (4.8%/year vs 0.8%/year; P |
---|---|
ISSN: | 1547-5271 1556-3871 |
DOI: | 10.1016/j.hrthm.2012.10.017 |