Comparison of 2 Portable Respiratory Gas Analysers
Abstract The aim of this study was to compare respiratory measures taken simultaneously using the Cosmed K4b 2 and Cortex Metamax II portable metabolic systems. 10 trained male cyclists performed a graded exercise cycle test to exhaustion (40w.3 min − 1 ) under standardized conditions. The measured...
Gespeichert in:
Veröffentlicht in: | International journal of sports medicine 2012-09, Vol.33 (9), p.728-733 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
The aim of this study was to compare respiratory measures taken simultaneously using the Cosmed K4b
2
and Cortex Metamax II portable metabolic systems. 10 trained male cyclists performed a graded exercise cycle test to exhaustion (40w.3 min
− 1
) under standardized conditions. The measured respiratory variables were significantly correlated between both devices: r=0.97 and 0.98 (n=10, p |
---|---|
ISSN: | 0172-4622 1439-3964 |
DOI: | 10.1055/s-0031-1301316 |