Unwanted phrenic nerve stimulation (PNS) has been reported in ∼1 in 4 patients undergoing left ventricular (LV) pacing. The occurrence of PNS over mid-term follow-up and the significance of PNS are less certain. Data from 1307 patients enrolled in pre-market studies of LV leads manufactured by Medtr...
Gespeichert in:
Veröffentlicht in: | Europace (London, England) England), 2013-01, Vol.15 (1), p.77-82 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Unwanted phrenic nerve stimulation (PNS) has been reported in ∼1 in 4 patients undergoing left ventricular (LV) pacing. The occurrence of PNS over mid-term follow-up and the significance of PNS are less certain.
Data from 1307 patients enrolled in pre-market studies of LV leads manufactured by Medtronic (models 4193 and 4195 unipolar, 4194, 4196, 4296, and 4396 bipolar) were pooled. Left ventricular lead location was recorded at implant using a common classification scheme. Phrenic nerve stimulation symptoms were either spontaneously reported or identified at scheduled follow-up visits. A PNS-related complication was defined as PNS resulting in invasive intervention or the termination of LV pacing. Average follow-up was 14.9 months (range 0.0-46.6). Phrenic nerve stimulation symptoms occurred in 169 patients (12.9%). Phrenic nerve stimulation-related complications occurred in 21 of 1307 patients (1.6%); 16 of 738 (2.2%) in the unipolar lead studies, and 5 of 569 (0.9%) in the bipolar lead studies (P = 0.08). Phrenic nerve stimulation was more frequent at middle-lateral/posterior, and apical LV sites (139/1010) vs. basal-posterior/lateral/anterior, and middle-anterior sites (20/297; P= 0.01). As compared with an anterior LV lead position, a lateral LV pacing site was associated with over a four-fold higher risk of PNS (P= 0.005) and an apical LV pacing site was associated with over six-fold higher risk of PNS (P= 0.001).
Phrenic nerve stimulation occurred in 13% of patients undergoing LV lead placement and was more common at mid-lateral/posterior, and LV apical sites. Most cases (123/139; 88%) of PNS were mitigated via electrical reprogramming, without the need for invasive intervention. |
---|---|