Determination of the energy response of closely tissue-equivalent diamond dosimeters for radiotherapy dosimetry
A monoenergetic X-ray syncrotron source was used to determine the energy response of a diamond detector in the range 5–25keV, clarifying the elemental composition of the detector. The response is shown to be influenced by the detector housing and electrical contacts. A model for the energy response...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2012-12, Vol.71, p.23-24 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A monoenergetic X-ray syncrotron source was used to determine the energy response of a diamond detector in the range 5–25keV, clarifying the elemental composition of the detector. The response is shown to be influenced by the detector housing and electrical contacts. A model for the energy response of the detector is determined that is valid in the 5keV–15MeV with an accuracy of 5% and therefore can be used to correct the dosimeter response to low-energy and scattered radiation.
► Response of diamond detector to monoenergetic synchrotron X-rays measured. ► Data from response at atomic orbital edges extends model of energy response. ► Corrects field dependent response of the detector due to scattered photons. |
---|---|