High Ventricular Lead Impedance of a DDD Pacemaker after Cranial Magnetic Resonance Imaging

Management of electromagnetic interference in the form of magnetic resonance imaging (MRI) in patients with pacemakers (PMs) may be challenging. Serious consequences, especially in PM‐dependent patients, may be encountered. Changes in device programming, asynchronous pacing, heating of the lead tip(...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2012-09, Vol.35 (9), p.e251-e253
Hauptverfasser: BASER, KAZIM, GURAY, UMIT, DURUKAN, MINE, DEMIRKAN, BURCU
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Management of electromagnetic interference in the form of magnetic resonance imaging (MRI) in patients with pacemakers (PMs) may be challenging. Serious consequences, especially in PM‐dependent patients, may be encountered. Changes in device programming, asynchronous pacing, heating of the lead tip(s), and increased thresholds or even device dislocation may be experienced. We report of a patient with a DDD PM who underwent an emergent MRI, after which there was an increase in ventricular impedance as well as increased cardiac biomarkers. (PACE 2012; 35:e251–e253)
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.2011.03027.x