Novel Sensors for Batch and Flow Injection Analysis of Histamine Based on Crown Ethers

Six types of histamine potentiometric sensors are developed. They are based on using dibenzo‐30‐crown‐10 (DB30C10) with potassium tetrakis(p‐chlorophenyl)borate lipophilic additive (Type I), dibenzo‐30‐crown‐10 without additive (Type II), dibenzo‐24‐crown‐8 (DB24C8) with the same additive (Type III)...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Electroanalysis (New York, N.Y.) N.Y.), 2012-05, Vol.24 (5), p.1226-1235
1. Verfasser: Elmosallamy, Mohamed A. F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Six types of histamine potentiometric sensors are developed. They are based on using dibenzo‐30‐crown‐10 (DB30C10) with potassium tetrakis(p‐chlorophenyl)borate lipophilic additive (Type I), dibenzo‐30‐crown‐10 without additive (Type II), dibenzo‐24‐crown‐8 (DB24C8) with the same additive (Type III), dibenzo‐24‐crown‐8 without additive (Type IV), dibenzo‐18‐crown‐6 with the additive (Type V) and dibenzo‐18‐crown‐6 without additive (Type VI) as neutral carriers for histamine. Sensors based on dibenzo‐30‐crown‐10 with (PTp‐C1PB) lipophilic additive (Type I) and dibenzo‐30‐crown‐10 without additive (Type II) show good response. The other sensors Types III–VI show poor response in terms of calibration range and slope.
ISSN:1040-0397
1521-4109
DOI:10.1002/elan.201100674