Rapid determination of ambrisentan enantiomers by enantioselective liquid chromatography using cellulose-based chiral stationary phase in reverse phase mode

A sensitive, specific, and rapid high‐performance liquid chromatography (HPLC) method for the determination of ambrisentan enantiomers has been developed and validated. Six chiral columns were tested in a reversed‐phase system. Excellent enantioseparation with the resolution more than 2.5 was achiev...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of separation science 2012-04, Vol.35 (7), p.798-803
Hauptverfasser: Douša, Michal, Gibala, Petr
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A sensitive, specific, and rapid high‐performance liquid chromatography (HPLC) method for the determination of ambrisentan enantiomers has been developed and validated. Six chiral columns were tested in a reversed‐phase system. Excellent enantioseparation with the resolution more than 2.5 was achieved on Chiralcel OZ‐3R (cellulose 3‐chloro‐4‐methylphenylcarbamate) using mixture of 20 mM sodium formate (pH 3.0) with acetonitrile (55:45; v/v). Validation of the HPLC method including linearity, limit of detection, limit of quantification, precision, accuracy, and selectivity was performed according to the International Conference on Harmonisation (ICH) guidelines. The method has an advantage of a very quick chromatographic separation (less than 6 min) and therefore is highly suitable for routine determination of (R)‐ambrisentan in enantiopure active pharmaceutical ingredient (S)‐ambrisentan.
ISSN:1615-9306
1615-9314
DOI:10.1002/jssc.201101062