Temperature Variation During Apicectomy with Er:YAG Laser
Objective: The purpose of this in vitro study was to evaluate the generated temperature of the Er:YAG laser, with three different pulse durations for apicectomy, compared with tungsten bur and surgical saw. Background data: Apicectomy is an endodontic surgery performed to remove the root apex and cu...
Gespeichert in:
Veröffentlicht in: | Photomedicine and laser surgery 2012-08, Vol.30 (8), p.425-428 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective:
The purpose of this
in vitro
study was to evaluate the generated temperature of the Er:YAG laser, with three different pulse durations for apicectomy, compared with tungsten bur and surgical saw.
Background data:
Apicectomy is an endodontic surgery performed to remove the root apex and curette adjacent periapical tissue because of lesions of the apical area that are not healing properly.
Methods:
Sixty single-rooted extracted human teeth were resected by three cutting methods: tungsten bur, surgical saw, and Er:YAG laser irradiation with three different pulse durations; pulse duration 50 μs, pulse duration 100 μs, and pulse duration 300 μs. Teflon-insulated, type K thermocouples were used to measure temperature changes during the apicectomy process. Data were analyzed using the general linear models procedure of the SPSS statistical software program.
Results:
Although there was no statistically significant difference for the mean values of temperature changes at 1 mm away to the cutting site of teeth, there was statistically significant difference among groups for the mean values of temperature changes at 3 mm away to the cutting site of teeth. Additionally, there was statistically significant difference among groups for the total time required for apicectomy.
Conclusions:
The laser irradiation with pulse duration 50 μs appears to have the lowest temperature rise and the shortest time required for apicectomy of the three pulse durations. However, Er:YAG laser for apicectomy in all pulse durations could be used safely for resection in endodontics in the presence of sufficient water. |
---|---|
ISSN: | 1549-5418 1557-8550 |
DOI: | 10.1089/pho.2012.3256 |