Measurements and detection of abdominal aortic aneurysm growth: Accuracy and reproducibility of a segmentation software

Abstract Purpose To validate the reproducibility and accuracy of a software dedicated to measure abdominal aortic aneurysm (AAA) diameter, volume and growth over time. Materials and methods A software enabling AAA segmentation, diameter and volume measurement on computed tomography angiography (CTA)...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of radiology 2012-08, Vol.81 (8), p.1688-1694
Hauptverfasser: Kauffmann, Claude, Tang, An, Therasse, Éric, Giroux, Marie-France, Elkouri, Stephane, Melanson, Philippe, Melanson, Bertrand, Oliva, Vincent L, Soulez, Gilles
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Purpose To validate the reproducibility and accuracy of a software dedicated to measure abdominal aortic aneurysm (AAA) diameter, volume and growth over time. Materials and methods A software enabling AAA segmentation, diameter and volume measurement on computed tomography angiography (CTA) was tested. Validation was conducted in 28 patients with an AAA having 2 consecutive CTA examinations. The segmentation was performed twice by a senior radiologist and once by 3 medical students on all 56 CTAs. Intra and inter-observer reproducibility of D -max and volumes values were calculated by intraclass correlation coefficient (ICC). Systematic errors were evaluated by Bland–Altman analysis. Differences in D -max and volume growth were compared with paired Student's t -tests. Results Mean D -max and volume were 49.6 ± 6.2 mm and 117.2 ± 36.2 ml for baseline and 53.6 ± 7.9 mm and 139.6 ± 56.3 ml for follow-up studies. Volume growth (17.3%) was higher than D -max progression (8.0%) between baseline and follow-up examinations ( p < .0001). For the senior radiologist, intra-observer ICC of D -max and volume measurements were respectively estimated at 0.997 (≥0.991) and 1.000 (≥0.999). Overall inter-observer ICC of D -max and volume measurements were respectively estimated at 0.995 (0.990–0.997) and 0.999 (>0.999). Bland–Altman analysis showed excellent inter-reader agreement with a repeatability coefficient
ISSN:0720-048X
1872-7727
DOI:10.1016/j.ejrad.2011.04.044