Noninvasive detection of mechanical prosthetic heart valve disorder
Abstract Auscultation is a widely used efficient technique by cardiologists for detecting the heart conditions. Since the mechanical prosthetic heart valves are widely used today, it is important to develop a simple and efficient method to detect abnormal mechanical valves. In this paper, the mechan...
Gespeichert in:
Veröffentlicht in: | Computers in biology and medicine 2012-08, Vol.42 (8), p.785-792 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Auscultation is a widely used efficient technique by cardiologists for detecting the heart conditions. Since the mechanical prosthetic heart valves are widely used today, it is important to develop a simple and efficient method to detect abnormal mechanical valves. In this paper, the mechanical prosthetic heart valve sounds are analyzed by using different power spectral density (PSD) estimation techniques. To improve the classification accuracy of heart sounds, we propose two different feature extraction schemes, i.e., a modified local discriminant bases (LDB) scheme and a Hilbert–Huang Transform (HHT)-based scheme. A database of 150 heart sounds is used in this study and an average classification accuracy of 97.3% is achieved for both the two feature extraction schemes, when a generic linear discriminant analysis (LDA) classifier is used in the classification stage. |
---|---|
ISSN: | 0010-4825 1879-0534 |
DOI: | 10.1016/j.compbiomed.2012.06.002 |