Assembly of an alpha–gamma coincidence measuring device for checking alpha decay schemes
Two new chambers for measuring alpha-particle emissions have been made: a low-geometry chamber with a powerful magnet to eliminate conversion electrons, and an alpha–gamma coincidence chamber. Both devices incorporate a high-resolution Si detector, and the second chamber, a low-energy Ge detector as...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2012-09, Vol.70 (9), p.2263-2266 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Two new chambers for measuring alpha-particle emissions have been made: a low-geometry chamber with a powerful magnet to eliminate conversion electrons, and an alpha–gamma coincidence chamber. Both devices incorporate a high-resolution Si detector, and the second chamber, a low-energy Ge detector as well. A dual parameter multichannel analyzer was used to register coincidences in the second device. Alpha-particle and gamma-ray detectors work simultaneously in both individual and dual modes, providing single and coincidence spectra. Some preliminary alpha–gamma coincidence spectra have been obtained.
► Two new chambers for measuring alpha-particle emissions have been made. ► Both devices have Si detectors and the second chamber a low-energy Ge detector also. ► A dual parameter multichannel analyzer was used to register coincidences. ► Alpha-particle and gamma-ray detectors work simultaneously. ► Some preliminary alpha–gamma coincidence spectra have been obtained. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2012.02.112 |