Performance evaluation of commercial radionuclide calibrators in Indonesian hospitals
Dose calibrator is one of the supporting equipments in the field of nuclear medicine. At the hospitals, dose calibrator is used for activity measurement of radiopharmaceutical before it is administered to patients. Comparison of activity measurements of 131I and 99mTc with dose calibrators was organ...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2012-09, Vol.70 (9), p.2243-2245 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Dose calibrator is one of the supporting equipments in the field of nuclear medicine. At the hospitals, dose calibrator is used for activity measurement of radiopharmaceutical before it is administered to patients. Comparison of activity measurements of 131I and 99mTc with dose calibrators was organized in Indonesia during 2007–2010 with the the aim of obtaining information dose calibrator performance in the hospitals. Seven Indonesian hospitals participated in this comparison. The measurement results were evaluated using the En criteria. The result presented in this paper facilitated the evaluation of dose calibrator performance at several hospitals.
► National comparisons of 131I and 99mTc radionuclides in Indonesian hospitals. ► Standardization using a Centronic IG11/A20 4πγ Ionization Chamber and participants using commercial radionuclide calibrators. ► Performance radionuclide calibrator in nuclear medicine in Indonesia. ► Measurement of activity of 99mTc and 131I was found satisfactory. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2012.02.113 |