Cerebral blood flow modulation by transcutaneous cranial electrical stimulation with Limoge's current
Summary Objectives Transcutaneous cranial electrical stimulation (TCES) delivers a high-frequency (166 kHz) pulsed biphasic balanced current with a pulse repetition frequency of 100 Hz with 40% duty cycle through a negative electrode and two positive electrodes over the skull. TCES has a proven abil...
Gespeichert in:
Veröffentlicht in: | Journal of neuroradiology 2012-07, Vol.39 (3), p.167-175 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Summary Objectives Transcutaneous cranial electrical stimulation (TCES) delivers a high-frequency (166 kHz) pulsed biphasic balanced current with a pulse repetition frequency of 100 Hz with 40% duty cycle through a negative electrode and two positive electrodes over the skull. TCES has a proven ability to potentiate anesthesia and analgesia, although the physiological mechanisms of this effect remain unclear. We hypothesized that the mechanism is a modulation of CBF in the central endogenous opioid system. This study aimed at determining the effects of TCES on CBF to elucidate its physiological mechanism. Methods Thirty-six healthy volunteers were randomly assigned to active or placebo TCES, and all assessments were double blind. TCES was performed using the Anesthelec™ device. In the stimulated group, an active cable was used, and in the control group (sham), the cable was inactive. CBF was measured by XeCT™ before and after two hours of TCES. Results Globally, CBF was unchanged by TCES. However, locally, TCES induced a significant CBF decrease in the brainstem and thalamus, which are structures involved in pain and anxiety (TCES and control CBF decrease were 18.5 and 11.9 mL/100 g brain tissue/min, respectively). Conclusion TCES can modulate local CBF but it has no effect on overall CBF. [Clinical Trials. gov number: NCT00273663 ]. |
---|---|
ISSN: | 0150-9861 |
DOI: | 10.1016/j.neurad.2011.06.001 |