Quantitative myocardial perfusion measurement using CT Perfusion: a validation study in a porcine model of reperfused acute myocardial infarction

We validated a CT Perfusion technique with beam hardening (BH) correction for quantitative measurement of myocardial blood flow (MBF). Acute myocardial infarction (AMI) was created in four pigs by occluding the distal LAD for 1 h followed by reperfusion. MBF was measured from dynamic contrast enhanc...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The International Journal of Cardiovascular Imaging 2012-06, Vol.28 (5), p.1237-1248
Hauptverfasser: So, Aaron, Hsieh, Jiang, Li, Jian-Ying, Hadway, Jennifer, Kong, Hua-Fu, Lee, Ting-Yim
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We validated a CT Perfusion technique with beam hardening (BH) correction for quantitative measurement of myocardial blood flow (MBF). Acute myocardial infarction (AMI) was created in four pigs by occluding the distal LAD for 1 h followed by reperfusion. MBF was measured from dynamic contrast enhanced CT (DCE-CT) scanning of the heart, with correction of cardiac motion and BH, before ischemic insult and on day 7, 10 and 14 post. On day 14 post, radiolabeled microspheres were injected to measure MBF and the results were compared with those measured by CT Perfusion. Excised hearts were stained with 2,3,5-triphenyltetrazolium chloride (TTC) to determine the relationship between MBF measured by CT Perfusion and myocardial viability. MBF measured by CT Perfusion was strongly correlated with that by microspheres over a wide range of MBF values (R = 0.81, from 25 to 225 ml min −1 100 g −1 ). While MBF in the LAD territory decreased significantly from 98.4 ± 2.5 ml min −1 100 g −1 at baseline to 32.2 ± 9.1 ml min −1 100 g −1 , P  
ISSN:1569-5794
1573-0743
1875-8312
DOI:10.1007/s10554-011-9927-x