CT Angiography of the Renal Arteries: Comparison of Lower-Tube-Voltage CTA With Moderate Concentration Iodinated Contrast Material and Conventional CTA

The objective of our study was to investigate the feasibility of an 80-kVp protocol using a moderate-concentration contrast medium (CM) for CT angiography of the renal arteries by comparison with a conventional 120-kVp protocol using a high-concentration CM. Attenuation values and signal-to-noise ra...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of roentgenology (1976) 2012-07, Vol.199 (1), p.96-102
Hauptverfasser: CHO, Eun-Suk, YU, Jeong-Sik, AHN, Jhii-Hyun, JOO HEE KIM, CHUNG, Jae-Joon, LEE, Hyeon-Kyeong, KYUNG HEE LEE
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The objective of our study was to investigate the feasibility of an 80-kVp protocol using a moderate-concentration contrast medium (CM) for CT angiography of the renal arteries by comparison with a conventional 120-kVp protocol using a high-concentration CM. Attenuation values and signal-to-noise ratios (SNRs) were determined in a phantom for the 120-kVp protocol with a high-concentration CM and the 80-kVp protocol with a moderate-concentration CM. In addition, 50 patients were prospectively enrolled in the study: 25 were scanned with 120 kVp and 200 effective mAs (mAs(eff)) after the administration of 110 mL of high-concentration CM (370 mg I/mL), and the other 25 were scanned with 80 kVp and 585 mAs(eff) after the administration of 110 mL of moderate-concentration CM (300 mg I/mL). Images of the two groups were compared in terms of arterial attenuation, SNR, contrast-to-noise ratio (CNR), and subjective degree of arterial enhancement and image quality. The mean attenuation of the main renal artery was significantly higher (p < 0.001) in the 80-kVp group who received moderate-concentration CM (mean ± SD, 370.0 ± 65.0 HU) than in the 120-kVp group who received high-concentration CM (269.9 ± 27.8 HU) without significant differences in SNR and CNR values. The 80-kVp protocol had significantly higher quality scores for arterial enhancement, sharpness of the artery, and overall diagnostic quality compared with the 120-kVp protocol. The effective dose of the 80-kVp protocol (4.5 mSv) was 8.2% lower than that of the 120-kVp protocol (4.9 mSv). The use of 80 kVp with moderate-concentration CM could improve arterial enhancement and provide superior image quality with a smaller amount of iodine and a lower radiation dose.
ISSN:0361-803X
1546-3141