Response and noise characteristics of small-sized inorganic and organic scintillation detectors measured with vacuum and solid-state photodetectors

Response and noise properties of scintillation detectors were investigated using a 137Cs radioactive source. The experimental setup consisted of a scintillator, photodetector, amplifier, and spectrometric analog-to-digital converter. Small-sized, cubic crystals were manufactured of stilbene, p-terph...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2012-02, Vol.664 (1), p.148-153
Hauptverfasser: Dudnik, O.V., Kurbatov, E.V., Tarasov, V.A., Andryushenko, L.A., Boyarintsev, A.Yu, Valtonen, E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Response and noise properties of scintillation detectors were investigated using a 137Cs radioactive source. The experimental setup consisted of a scintillator, photodetector, amplifier, and spectrometric analog-to-digital converter. Small-sized, cubic crystals were manufactured of stilbene, p-terphenyl and thallium-doped cesium iodide, as well as of a plastic scintillator material based on polystyrene. Vacuum and solid-state photomultipliers and a silicon PIN photodiode were used as photodetectors. The energy equivalent noise was determined for all used photodetectors. The response of various scintillators when using vacuum and semiconductor photomultipliers was found to be closely similar.
ISSN:0168-9002
1872-9576
DOI:10.1016/j.nima.2011.10.051