A Biomechanical Comparison of a Locking Plate, a Nail, and a 95° Angled Blade Plate for Fixation of Subtrochanteric Femoral Fractures

OBJECTIVE:To compare the biomechanical performance of a cephalomedullary nail (CMN), a proximal femoral locking plate, and a 95° angled blade plate in a comminuted subtrochanteric fracture model. METHODS:A comminuted subtrochanteric femoral fracture model was created with a 2-cm gap below the lesser...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of orthopaedic trauma 2012-06, Vol.26 (6), p.334-340
Hauptverfasser: Forward, Daren P, Doro, Christopher J, OʼToole, Robert V, Kim, Hyunchul, Floyd, John C P, Sciadini, Marcus F, Turen, Clifford H, Hsieh, Adam H, Nascone, Jason W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:OBJECTIVE:To compare the biomechanical performance of a cephalomedullary nail (CMN), a proximal femoral locking plate, and a 95° angled blade plate in a comminuted subtrochanteric fracture model. METHODS:A comminuted subtrochanteric femoral fracture model was created with a 2-cm gap below the lesser trochanter in 15 pairs of human cadaveric femora confirmed to be nonosteoporotic. The femora were randomized to treatment with one of the previously mentioned 3 devices. Each was tested under incrementally increasing cyclic load up to 90,000 cycles from 50% to 250% of body weight to simulate progressive weight bearing during 3 months of an average 700-N (approximately, 70 kg or 150 lb) person. Force, number of cycles, and total load sustained to reach 10 mm of displacement were compared. Failure modes were also noted. RESULTS:The CMN construct withstood significantly more cycles, failed at a significantly higher force, and withstood a significantly greater load than either of the plate constructs (P < 0.001). Varus collapse was significantly lower in the CMN construct (P < 0.0001). Modes of failure differed among implant–bone constructs with damage to the femoral head through implant cutout in 5 of 10 blade plate specimens and 2 of 10 CMN specimens, whereas no damage to the femoral head bone was observed in any of the locking plate constructs. CONCLUSIONS:The CMN construct was biomechanically superior to either the locking plate or 95° blade plate constructs. The locking plate construct was biomechanically equivalent to the blade plate construct.
ISSN:0890-5339
1531-2291
DOI:10.1097/BOT.0b013e3182254ea3