Variability of the minimum detectable energy of an APD as an electron detector

The purpose of this study is to evaluate the variability of the minimum detectable energy of the avalanche photodiode which would be applied for measurements of medium-energy ( ∼ 10 – 100 keV ) electrons in future magnetospheric exploration. The minimum detectable energy is affected by temperature v...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2012-02, Vol.664 (1), p.282-288
Hauptverfasser: Kasahara, S., Takashima, T., Hirahara, M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study is to evaluate the variability of the minimum detectable energy of the avalanche photodiode which would be applied for measurements of medium-energy ( ∼ 10 – 100 keV ) electrons in future magnetospheric exploration. The minimum detectable energy is affected by temperature variations as well as radiation. Our experiments and theoretical considerations show that the minimum detectable energy is significantly deteriorated at room temperature after proton irradiation of > 10 krad mainly due to an increase in the dark current, whereas such a degradation is less prominent when the device is cooled down to < 5 ° C . These results provide a guideline for the thermal design of the medium-energy electron instruments especially for harsh-radiation space missions.
ISSN:0168-9002
1872-9576
DOI:10.1016/j.nima.2011.11.033