Fe–C eutectic fixed-point cells for contact thermometry: an investigation and comparison
Five iron-carbon (Fe-C) eutectic fixed-point cells have been constructed between NPL and LNE-Cnam to investigate the robustness and to measure the agreement of their melting temperatures. Each cell was constructed with a different selection of materials sourced by NPL and LNE-Cnam. The measured emfs...
Gespeichert in:
Veröffentlicht in: | Metrologia 2012-02, Vol.49 (1), p.88-94 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Five iron-carbon (Fe-C) eutectic fixed-point cells have been constructed between NPL and LNE-Cnam to investigate the robustness and to measure the agreement of their melting temperatures. Each cell was constructed with a different selection of materials sourced by NPL and LNE-Cnam. The measured emfs at the Fe-C fixed-point temperature (~1153 ...C), compared between cells, agree within around 1.98 ...V (~90 mK), where the most important contribution to the uncertainty of each measurement is the inhomogeneity associated with the measuring Pt/Pd thermocouple. This demonstrates that these cells are suitable for use as secondary fixed-point cells in contact thermometry but the robustness of the presented cells is not found to be sufficient for maintaining their integrity during repeated cycling procedures. (ProQuest: ... denotes formulae/symbols omitted.) |
---|---|
ISSN: | 0026-1394 1681-7575 |
DOI: | 10.1088/0026-1394/49/1/013 |