Study of standard thermometers and their reproduction of the [lambda] point in liquid helium
Self-calibrated resistance thermometers to be used as standards have been investigated. The heat exchange of the sensor coil with liquid helium per unit coil surface area at 0.3 K is 12 times better than that of a conventional gas-filled thermometer. The exchange is due to Kapitza surface thermal re...
Gespeichert in:
Veröffentlicht in: | Measurement techniques 1999-07, Vol.42 (7), p.689 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Self-calibrated resistance thermometers to be used as standards have been investigated. The heat exchange of the sensor coil with liquid helium per unit coil surface area at 0.3 K is 12 times better than that of a conventional gas-filled thermometer. The exchange is due to Kapitza surface thermal resistance. The λ point in liquid helium filling of the thermometer is reproducible to within 0.1 mK.[PUBLICATION ABSTRACT] |
---|---|
ISSN: | 0543-1972 1573-8906 |
DOI: | 10.1007/BF02512093 |