Reducing the Detection Threshold in Electrothermal Atomic-Absorption Spectrometry
The influence of the process of signal filtration on the shape and noise component of analytic signals in electrothermal atomic-absorption spectrometry at rates of heating of graphite tube furnaces on the order of 10,000 K·sec^sup -1^ is investigated. A digital filter with finite impulse response an...
Gespeichert in:
Veröffentlicht in: | Measurement techniques 2001-08, Vol.44 (8), p.821 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The influence of the process of signal filtration on the shape and noise component of analytic signals in electrothermal atomic-absorption spectrometry at rates of heating of graphite tube furnaces on the order of 10,000 K·sec^sup -1^ is investigated. A digital filter with finite impulse response and symmetric coefficients by means of which the noise component of the signals can be reduced by 90% without any distortion in the signal shape is synthesized. This also helps to realize a 90% reduction in the detection threshold of elements.[PUBLICATION ABSTRACT] |
---|---|
ISSN: | 0543-1972 1573-8906 |
DOI: | 10.1023/A:1012996301569 |