Influence of Conversion Nonlinearity of a Bolometer on the Results of Measuring Nonuniformity of the Intensity Distribution of Laser Radiation
A derivation is given of the dependence of the nonuniformity coefficient of the intensity distribution of laser radiation, measured using a profile bolometer transducer, on the nonlinearity of the bolometer conversion characteristic and the effective power level or effective specific energy. The ran...
Gespeichert in:
Veröffentlicht in: | Measurement techniques 2001-06, Vol.44 (6), p.569 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A derivation is given of the dependence of the nonuniformity coefficient of the intensity distribution of laser radiation, measured using a profile bolometer transducer, on the nonlinearity of the bolometer conversion characteristic and the effective power level or effective specific energy. The random error of the measured coefficient is estimated.[PUBLICATION ABSTRACT] |
---|---|